FDA recall record

Microgenics Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control recall — FDA Z-0415-2026

The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect…

Class IIclassification
Z-0415-2026official record ID
Nov 5, 2025report date

Recall record at a glance

Official recordZ-0415-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMicrogenics Corporation
Report dateNov 5, 2025
Recall initiation dateOct 1, 2025
Quantity4462
LocationFremont, CA, United States

Official product description

Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control

Why the product was recalled

The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.

Distribution information

Worldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI, IL, LA, AR, TX, UT, GA, ID and the countries of MEX, CAN, KOR, MYS, DEU, GBR, BRA, JAP, HKG, NZL, AUS.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.