FDA recall record

SIGHT DIAGNOSTICS The Sight OLO device is a computer vision based platform for blood recall — FDA Z-0963-2022

In instances where custom reference ranges were configured on the device post installation, a possibility to inadvertently apply changes to reference…

Class IIclassification
Z-0963-2022official record ID
Apr 27, 2022report date

Recall record at a glance

Official recordZ-0963-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSIGHT DIAGNOSTICS LTD
Report dateApr 27, 2022
Recall initiation dateJan 20, 2022
Quantity322 units: U.S 109 Foreign 213
LocationTel Aviv-Yafo, Israel

Official product description

The Sight OLO device is a computer vision based platform for blood analysis. The platform combines computer-vision algorithms for image processing to identify and quantify blood components (e.g., red blood cells) and their characteristics (e.g., cell volume) in an automated fashion. Using dedicated staining, the proposed platform provides complete blood count analysis. The Sight OLO is a compact device, designed to be automated and simple to operate, to enable rapid testing and analysis. The Sight OLO consists of a scanning and analyzing device and a CBC test kit, including disposable cartridges and sample preparation tools. The disposable cartridge containing the blood sample is loaded into the device through the loading slot. The device is operated through the touch screen interface. The Sight OLO provides complete blood count information with 5-part differentials for white blood cell types. Specifically, the CBC parameters measured by the Sight OLO are listed below and include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, NEUT%/#, LYMPH %/#, MONO %/#, EO%/# and BASO%/#. In addition, the Sight OLO signals specific WBC abnormal cases by flagging the sample.

Why the product was recalled

In instances where custom reference ranges were configured on the device post installation, a possibility to inadvertently apply changes to reference range values was found on software versions 2.59.3 and all earlier versions, which can potentially lead to displaying and printing incorrectly-configured reference range values

Distribution information

Worldwide distribution - US Nationwide and the countries of AUSTRALIA, CANADA, CHILE, CURACAO, FINLAND, GEORGIA, GERMANY, GREECE, HONG KONG, IRELAND, ISRAEL, ITALY, MEXICO, NETHERLANDS, PORTUGAL, SINGAPORE, SLOVENIA, UNITED ARAB EMIRATES, UNITED KINGDOM.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 3, 2024

Sight OLO CBC Test Kit recall — FDA Z-2173-2024

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test…

Class IIFDA: OngoingSIGHT DIAGNOSTICS LTD