Sight OLO, Automated Hematology Analyzer, Catalog No recall — FDA Z-1469-2022

The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y" are mixed up and incorrectly displayed on OLO devices running software…

Class IIclassification
Z-1469-2022official record ID
Aug 3, 2022report date

Recall record at a glance

Official recordZ-1469-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSIGHT DIAGNOSTICS LTD
Report dateAug 3, 2022
Recall initiation dateJun 14, 2022
Quantity72 US; 292 OUS
LocationTel Aviv-Yafo, Israel

Official product description

Sight OLO, Automated Hematology Analyzer, Catalog No. OLO-U1

Why the product was recalled

The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y" are mixed up and incorrectly displayed on OLO devices running software version 2.63, 2.63.1, and 2.63.1.1. Due to the issue, results may incorrectly appear to be within or outside the reference range.

Distribution information

Worldwide distribution - US Nationwide distribution in the states of Alabama, Arizona, California, Florida, Idaho, Illinois, Kansas, Kentucky, Massachusetts, Michigan, Minnesota, Mississippi, New Jersey, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin and the countries of Burkina Faso, Belize, Canada, Cote D'Ivoire, Cameroon, Curacao, Germany, Spain, Finland, United Kingdom, Greece, Hong Kong, Indonesia, Ireland, Israel, India, Italy, Mexico, Portugal, Qatar, Sweden, Slovenia, Chad, Thailand, South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 3, 2024

Sight OLO CBC Test Kit recall — FDA Z-2173-2024

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test…

Class IIFDA: OngoingSIGHT DIAGNOSTICS LTD