FDA recall record

Merge Healthcare The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a recall — FDA Z-2707-2017

If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the…

Class IIclassification
Z-2707-2017official record ID
Jul 19, 2017report date

Recall record at a glance

Official recordZ-2707-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMerge Healthcare, Inc.
Report dateJul 19, 2017
Recall initiation dateMar 10, 2016
Quantity555 PDM's
LocationHartland, WI, United States

Official product description

The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.

Why the product was recalled

If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.

Distribution information

Distribution US nationwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.