Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 recall — FDA Z-2518-2025

Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain…

Class IIclassification
Z-2518-2025official record ID
Sep 10, 2025report date

Recall record at a glance

Official recordZ-2518-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMerge Healthcare, Inc.
Report dateSep 10, 2025
Recall initiation dateAug 11, 2025
Quantity84
LocationHartland, WI, United States

Official product description

Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedures. The procedures can be performed in the clinical areas of invasive cardiology, interventional radiology, and cardiac electrophysiology.

Why the product was recalled

Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.

Distribution information

US Nationwide distribution including in the states of Arizona, Arkansas, California, Florida, Louisiana, Mississippi, North Carolina, North Dakota, Oklahoma, Pennsylvania, Georgia, Illinois, Kentucky, Tennessee, Texas, Washington, and Wisconsin.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.