FDA recall record

BioMerieux SA The PREVI Color Gram dyes are used with the PREVI Color Gram instrument to recall — FDA Z-2170-2018

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply…

Class IIclassification
Z-2170-2018official record ID
Jun 20, 2018report date

Recall record at a glance

Official recordZ-2170-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBioMerieux SA
Report dateJun 20, 2018
Recall initiation dateMar 23, 2018
QuantityFUCHSIN-A (ref: 29522) 15 eaches; IODINE-B (ref: 29523) 10 eaches ; CRYSTAL VIOLET-C (ref: 29524) 10 eaches
LocationMarcy L'Etoile, N/A, France

Official product description

The PREVI Color Gram dyes are used with the PREVI Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram s method. The PREVI Color Gram instrument sprays the PREVI Color Gram reagents onto microscope slides which have been smeared with specimens to be examined. The staining procedure is rapid and precise. Each reagent has a specific position (A, B, C, D, E) on the PREVI Color Gram platform that is also indicated on the reagent label. An automated slide stainer is a device used to stain histology, cytology, and hematology slides for diagnosis.

Why the product was recalled

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Distribution information

International distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.