FDA recall record

BioMerieux SA VIDAS TOXO IgG Avidity is an automated qualitative test for use on the VIDAS recall — FDA Z-2169-2018

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply…

Class IIclassification
Z-2169-2018official record ID
Jun 20, 2018report date

Recall record at a glance

Official recordZ-2169-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBioMerieux SA
Report dateJun 20, 2018
Recall initiation dateMar 23, 2018
Quantity2 boxes VIDAS TOXO IGG AVIDITY 30 TESTS
LocationMarcy L'Etoile, N/A, France

Official product description

VIDAS TOXO IgG Avidity is an automated qualitative test for use on the VIDAS family instruments, for the determination of anti-toxoplasma IgG avidity in human serum or plasma (lithium heparin, sodium citrate, EDTA) using the ELFA technique (Enzyme Linked Fluorescent Assay). Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens.

Why the product was recalled

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Distribution information

International distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.