FDA recall record

Medacta Usa The GMK knee system: designed for cemented use in total knee arthroplasty recall — FDA Z-0266-2018

Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance…

Class IIclassification
Z-0266-2018official record ID
Dec 27, 2017report date

Recall record at a glance

Official recordZ-0266-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedacta Usa Inc.
Report dateDec 27, 2017
Recall initiation dateOct 24, 2017
Quantity12,768 units
LocationMemphis, TN, United States

Official product description

The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF

Why the product was recalled

Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.

Distribution information

nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.