FDA recall record

Medacta Usa Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971 recall — FDA Z-2618-2026

Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and…

Class IIclassification
Z-2618-2026official record ID
Jul 8, 2026report date

Recall record at a glance

Official recordZ-2618-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedacta Usa Inc
Report dateJul 8, 2026
Recall initiation dateApr 30, 2026
Quantity50
LocationMemphis, TN, United States

Official product description

Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System

Why the product was recalled

Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.

Distribution information

US Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: N/A.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.