FDA recall record

Insulet The failure occurs if a user on the Omnipod 5 iOS application selects and holds down recall — FDA Z-0132-2026

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the…

Class IIclassification
Z-0132-2026official record ID
Oct 22, 2025report date

Recall record at a glance

Official recordZ-0132-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmInsulet Corporation
Report dateOct 22, 2025
Recall initiation dateSep 11, 2025
Quantity174,013 units
LocationActon, MA, United States

Official product description

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

Why the product was recalled

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 8, 2026

Insulet Omnipod (Eros) Pod recall — FDA Z-2605-2026

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary…

Class IFDA: OngoingInsulet Corporation
Medical DeviceApr 22, 2026

Insulet Omnipod 5 Pods recall — FDA Z-1797-2026

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user…

Class IFDA: OngoingInsulet Corporation
Medical DeviceJan 10, 2024

Insulet Omnipod 5 App recall — FDA Z-0591-2024

The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user…

Class IFDA: OngoingInsulet Corporation