Recall record at a glance
| Official record | Z-0591-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Ongoing |
| Recalling firm | Insulet Corporation |
| Report date | Jan 10, 2024 |
| Recall initiation date | Nov 30, 2023 |
| Quantity | 27000 users |
| Location | Acton, MA, United States |
Official product description
Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.
Why the product was recalled
The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.
Distribution information
Worldwide distribution - US Nationwide and the countries of Germany and UK.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.