FDA recall record

Maquet Medical Systems USA TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet recall — FDA Z-0993-2015

The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the…

Class IIclassification
Z-0993-2015official record ID
Jan 28, 2015report date

Recall record at a glance

Official recordZ-0993-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMaquet Medical Systems USA
Report dateJan 28, 2015
Recall initiation dateAug 21, 2014
Quantity6 units
LocationWayne, NJ, United States

Official product description

TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet Tegris OR Integration System is designed to be used as the central operating unit in an operating room. The integration system has two main functions: Recording and distribution of images and videos and interaction with medical and non-medical devices.

Why the product was recalled

The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.

Distribution information

US Nationwide Distribution including NE, NY and PA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.