Maquet Medical Systems USA Venous probe connection cable recall — FDA Z-0605-2024

Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may…

Class IIclassification
Z-0605-2024official record ID
Dec 27, 2023report date

Recall record at a glance

Official recordZ-0605-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMaquet Medical Systems USA
Report dateDec 27, 2023
Recall initiation dateNov 13, 2023
Quantity1213 US, 2589 OUS
LocationWayne, NJ, United States

Official product description

Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 and 701041802

Why the product was recalled

Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.

Distribution information

Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Islamic Republic Of Iraq, Italy, Japan, Kazakhstan, Korea, Republic Of Kuwait, Latvia, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Martinique, Mauritius, Mayotte, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.