FDA recall record

FUJIFILM Healthcare Americas Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0 recall — FDA Z-1171-2024

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main…

Class IIclassification
Z-1171-2024official record ID
Feb 28, 2024report date

Recall record at a glance

Official recordZ-1171-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmFUJIFILM Healthcare Americas Corporation
Report dateFeb 28, 2024
Recall initiation dateJan 3, 2024
Quantity13
LocationLexington, MA, United States

Official product description

Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.

Why the product was recalled

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Distribution information

US domestic distribution nationwide. International distribution to Israel.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.