FDA recall record

FUJIFILM Healthcare Americas FDR Visionary Suite recall — FDA Z-1299-2026

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current…

Class IIclassification
Z-1299-2026official record ID
Feb 11, 2026report date

Recall record at a glance

Official recordZ-1299-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmFUJIFILM Healthcare Americas Corporation
Report dateFeb 11, 2026
Recall initiation dateJan 9, 2026
Quantity16 units
LocationLexington, MA, United States

Official product description

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Why the product was recalled

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Distribution information

US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.