FDA recall record

Medtronic Navigation StealthStation Cranial Software, Models: 9735585, 9735586 (kit) recall — FDA Z-1630-2023

During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays…

Class Iclassification
Z-1630-2023official record ID
Jun 14, 2023report date

Recall record at a glance

Official recordZ-1630-2023
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmMedtronic Navigation, Inc.
Report dateJun 14, 2023
Recall initiation dateApr 11, 2023
Quantity746
LocationLafayette, CO, United States

Official product description

StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems

Why the product was recalled

During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury

Distribution information

Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA, Australia, AUSTRIA, BANGLADESH, BARBADOS, BELGIUM, Bosnia and Herzegovina, BULGARIA, CANADA, CHILE, CROATIA, CZECH REPUBLIC, DENMARK, FRANCE, GREECE, INDIA, IRELAND, ITALY, JAPAN, LUXEMBOURG, MARTINIQUE, NEW ZEALAND, POLAND, PORTUGAL, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVAKIA, SPAIN, SWITZERLAND, TAIWAN, Turkey, UNITED ARAB EMIRATES, UNITED KINGDOM

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.