FDA recall record

Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE recall — FDA Z-1022-2025

Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.

Class IIclassification
Z-1022-2025official record ID
Feb 5, 2025report date

Recall record at a glance

Official recordZ-1022-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedtronic Navigation, Inc.
Report dateFeb 5, 2025
Recall initiation dateDec 4, 2024
Quantity67,370 pins
LocationLafayette, CO, United States

Official product description

Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: 9733235 Medtronic PIN, 150MM, STERILE PERCUTANEOUS REF: 97733236 The Sterile Percutaneous Reference Pin Set is intended to be used to place and remove a pin percutaneously into bony patient anatomy for rigid attachment of a patient reference during image-guided surgeries using a Medtronic computer-assisted surgery system.

Why the product was recalled

Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.

Distribution information

Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Czech Republic, Denmark, Finland, Germany, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, United Arab Emirates, United Kingdom, Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.