Recall record at a glance
| Official record | Z-0417-2022 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Medtronic Navigation, Inc. |
| Report date | Dec 29, 2021 |
| Recall initiation date | Nov 9, 2021 |
| Quantity | 98 |
| Location | Louisville, CO, United States |
Official product description
Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.
Why the product was recalled
A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.
Distribution information
US: FL, NE, CA, MN, TX, OH, IL, AZ, KS, KY, OR, OK, WA, NV, MO, IN, MI, UT, MS, VA, SC. OUS: Canada, Sweden, Italy, France, Germany, Spain, Czech Republic, United Kingdom, India, Nepal, Japan, Turkey, China
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.