FDA recall record

Baxter Healthcare SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009 recall — FDA Z-0107-2019

If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator…

Class IIclassification
Z-0107-2019official record ID
Oct 24, 2018report date

Recall record at a glance

Official recordZ-0107-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBaxter Healthcare Corporation
Report dateOct 24, 2018
Recall initiation dateSep 18, 2018
Quantity354 units
LocationMedina, NY, United States

Official product description

SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009

Why the product was recalled

If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.

Distribution information

NJ

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.