FDA recall record

Baxter Healthcare SIGMA Spectrum Infusion Pump recall — FDA Z-1121-2026

Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the…

Class IIclassification
Z-1121-2026official record ID
Jan 28, 2026report date

Recall record at a glance

Official recordZ-1121-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBaxter Healthcare Corporation
Report dateJan 28, 2026
Recall initiation dateNov 28, 2025
Quantity585 units
LocationDeerfield, IL, United States

Official product description

SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2

Why the product was recalled

Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Exacta Mix Vented Micro-Volume Inlet recall — FDA Z-2262-2026

Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower…

Class IIFDA: OngoingBaxter Healthcare Corporation