Howmedica Osteonics Short tibial bearing MK4 - Std, STM, Catalog No recall — FDA Z-1778-2022

A MK4 Tibial Bearing was supplied with the intention to mate with the Tibial Metal Casing in-situ. However, the device in-situ was a MK3 Tibial Metal…

Class IIclassification
Z-1778-2022official record ID
Oct 5, 2022report date

Recall record at a glance

Official recordZ-1778-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmHowmedica Osteonics Corp.
Report dateOct 5, 2022
Recall initiation dateAug 24, 2022
Quantity1 unit
LocationMahwah, NJ, United States

Official product description

Short tibial bearing MK4 - Std, STM, Catalog No. SMMLTB02. Knee prosthesis component.

Why the product was recalled

A MK4 Tibial Bearing was supplied with the intention to mate with the Tibial Metal Casing in-situ. However, the device in-situ was a MK3 Tibial Metal Casing. The supplied MK4 tibial bearing was not exchangeable for the MK3 tibial bearing. The surgeon needed to complete the surgery by re-assembling and re-inserting the extracted MK3 tibial bearing.

Distribution information

Distributed to one customer located in New York.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.