FDA recall record

Howmedica Osteonics OMNIFIT SERIES I ACET recall — FDA Z-3196-2026

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the…

Class IIclassification
Z-3196-2026official record ID
Sep 23, 2026report date

Recall record at a glance

Official recordZ-3196-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmHowmedica Osteonics Corp.
Report dateSep 23, 2026
Recall initiation dateJul 30, 2026
Quantity12,156 US
LocationMahwah, NJ, United States

Official product description

OMNIFIT SERIES I ACET. INSERT 10 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Why the product was recalled

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Distribution information

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.