Intuitive Surgical SEAL,IS4000 PORTS,5MM-8MM,BOX of 10 recall — FDA Z-1836-2016

Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the…

Class IIclassification
Z-1836-2016official record ID
Jun 8, 2016report date

Recall record at a glance

Official recordZ-1836-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntuitive Surgical, Inc.
Report dateJun 8, 2016
Recall initiation dateMay 9, 2016
Quantity46, 191 boxes - total both sizes
LocationSunnyvale, CA, United States

Official product description

SEAL,IS4000 PORTS,5MM-8MM,BOX of 10; Model numbers: 470361-03, 470361-04, 470361-05, 470361-06; Part number: 470361. The da Vinci Xi Surgical System cannula sterile single use disposable cap for each cannula . When attached to the cannula, the seal maintains insufflation when the cannula is empty and when the instrument is inserted in the patient during a surgical procedure.

Why the product was recalled

Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula lumen when insufflation is connected during a surgical procedure.

Distribution information

Worldwide distribution. US, Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Da Vinci 5 ASSY,DV5 TOWER,IS5000 recall — FDA Z-2310-2026

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the…

Class IIFDA: OngoingIntuitive Surgical, Inc.