Pega Medical SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the recall — FDA Z-0324-2016

SCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.

Class IIclassification
Z-0324-2016official record ID
Dec 2, 2015report date

Recall record at a glance

Official recordZ-0324-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPega Medical Inc.
Report dateDec 2, 2015
Recall initiation dateOct 20, 2015
Quantity4 units distributed in the US
LocationLaval, N/A, Canada

Official product description

SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System. This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number. The Free-Gliding SCFE Screw System is a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients. Orthopedic.

Why the product was recalled

SCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.

Distribution information

Distributed in the states of FL and TN.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 13, 2019

Pega Medical SCFE recall — FDA Z-0270-2020

There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.

Class IIFDA: TerminatedPega Medical Inc.