FDA recall record

Pega Medical GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM recall — FDA Z-1362-2015

Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred…

Class IIclassification
Z-1362-2015official record ID
Apr 8, 2015report date

Recall record at a glance

Official recordZ-1362-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPega Medical Inc.
Report dateApr 8, 2015
Recall initiation dateFeb 20, 2015
Quantity7 units distributed in the US;
LocationLaval, N/A, Canada

Official product description

GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP Endo-Exo Medullary System is used for the treatment of fractures or correction of deformities in the femur, tibia and humerus of pediatric patients, who have reached skeletal maturity. The GAP Nail is an intramedullary nail. It is the main component of the GAP Endo-Exo Medullary System. The rest of the components of the system are screws, plates and caps that are assembled to the nail during the surgical procedure. The GAP Endo-Exo Medullary System is indicated as a temporary implant to assure alignment, stabilization and fixation of: long bones that have been surgically prepared (osteotomy) for correction of deformities or fractures caused by trauma or disease. The GAP System is used for pediatric patients (child and adolescent) ages 2 to 21. It can be used to correct the following conditions: -Diaphyseal fracture of the femur, tibia and humerus -Fractures of the femoral neck -Subtrochanteric, intertrochanteric and combination fractures -Correction of deformities (OI, Coxa vara, Coxa valga) -Nonunions and malunions

Why the product was recalled

Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred in patients with tibial pseudoarthrosis (CPT) or Osteogenesis imperfecta (OI). Higher failure rates have been detected in tibias with distal fractures that are close to the distal locking holes.

Distribution information

US Nationwide Distribution in the states of: California, Florida, New York, and Puerto Rico (US).

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 13, 2019

Pega Medical SCFE recall — FDA Z-0270-2020

There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.

Class IIFDA: TerminatedPega Medical Inc.