FDA recall record

NeuroLogica Samsung GM60A-40S Intended for use in generating radiographic images of human recall — FDA Z-0263-2016

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be…

Class IIclassification
Z-0263-2016official record ID
Nov 18, 2015report date

Recall record at a glance

Official recordZ-0263-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmNeuroLogica Corporation
Report dateNov 18, 2015
Recall initiation dateSep 14, 2015
Quantity7 devices
LocationDanvers, MA, United States

Official product description

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Why the product was recalled

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Distribution information

Distributed to the states of NC, OH, MA, and MO.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.