NeuroLogica GM85 Digital Mobile X-ray imaging System recall — FDA Z-0480-2026
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
Overview of recall records associated with NeuroLogica Corporation, including dates, categories and classifications where available.
Classification mix: Class II: 9.
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
Potential for the anti-fall system of the device arm to fail.
Mobile x-ray systems moving arm frame welding issue which presents the potential of bodily harm due to fall of the arm.
1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the…
There are potential issues related to the operational state of the scanner, specifically for transport and inadvertent system shutdowns. There could be a possibility of potential delay in…
There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.
There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.
Potential issue related to the operation of the safety latch within the column that supports the arm due to a possibility of potential bodily harm due to a fall of the arm in an unlikely…
While transporting the portable machine with the retrofit equipment in a backwards motion, there is potential for foot injury from the underside of the detector bin of the retrofit unit.
This page summarizes recall records in which the official source identifies NeuroLogica Corporation as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.