Beckman Coulter Mishima K.K Sample probe sucks a sample dispensed into a tube or cup and recall — FDA Z-1908-2025

Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be…

Class IIclassification
Z-1908-2025official record ID
Jun 11, 2025report date

Recall record at a glance

Official recordZ-1908-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusCompleted
Recalling firmBeckman Coulter Mishima K.K.
Report dateJun 11, 2025
Recall initiation dateApr 30, 2025
Quantity6,166
LocationSunto-Gun, N/A, Japan

Official product description

Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670

Why the product was recalled

Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.

Distribution information

Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX, NJ, FL, IN, MA, NY, NC, OH, LA, AZ, MO, WA, GA, UT, KY, AL, CO, IL, NE, OK, NV, SC, TN, MD, OR, AR, MS, AK, WI, DE, MT, HI, KS, ME, NH, WV, NM, CT, IA, MI, MN, ID, RI, SD, WY, ND, DC, VT, GUAM, PR and OUS (foreign) to countries of: Viet Nam, Japan, India, Australia, El Salvador, China, United Arab Emirates, Thailand, Russian Federation, Italy, Germany, Korea (Republic of, Malaysia), Taiwan, Province of China, Turkey, Kazakhstan, Israel, Ecuador, Hungary, Egypt, Serbia, Bangladesh, Bosnia and Herzegovina, Costa Rica, Belarus, United Kingdom, Ireland, Botswana, Kenya, Mauritius, Ghana, Zambia, South Africa, Libya, Croatia, France, Switzerland, Czechia, Philippines, Namibia, Canada, Qatar, Azerbaijan, Lesotho, Bulgaria, Zimbabwe, Poland, Norway, Spain, Brazil, Tanzania (United Republic of), Lithuania, Yemen, Uzbekistan, Algeria, Somalia, Sri Lanka, Nepal, Pakistan, Ukraine, Peru, Morocco, Singapore, Slovakia, Mexico, Myanmar, Lebanon, Jordan, Malawi, Greece, Indonesia, Chile, Romania, Georgia, Ethiopia, Sweden, Albania, Eswatini, Nigeria, Cote d'Ivoire, Argentina, Dominican Republic, Iraq, Maldives, Saudi Arabia, Kuwait, Austria, Eritrea, Bolivia (Plurinational State of), Tunisia, Curacao, Mozambique, Hong Kong, Bahrain, Portugal, Panama, New Zealand, Colombia, Uganda, Oman, Belgium, Saint Pierre and Miquelon, Guyana, Macao, Cabo Verde, Sint Maarten (Dutch part), Liberia, Moldova, Republic of, North Macedonia, Congo (Democratic Republic of the), Uruguay, Cyprus, Netherlands, Suriname, Slovenia, Denmark, Finland, Venezuela, Sierra Leone

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.