Beckman Coulter Mishima K.K DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an recall — FDA Z-2455-2024

There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed…

Class IIclassification
Z-2455-2024official record ID
Aug 7, 2024report date

Recall record at a glance

Official recordZ-2455-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter Mishima K.K.
Report dateAug 7, 2024
Recall initiation dateJun 5, 2024
Quantity24 analyzers
LocationSunto-Gun, N/A, Japan

Official product description

DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Why the product was recalled

There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. Issue 2: calibration with expired calibrator. They both may cause erroneous patient results, but the probability of serious adverse health consequences or death is unlikely.

Distribution information

Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, and South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.