Recall record at a glance
| Official record | Z-1280-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Baxter Healthcare Corporation |
| Report date | Apr 11, 2018 |
| Recall initiation date | Feb 15, 2017 |
| Quantity | 3,440 units |
| Location | Deerfield, IL, United States |
Official product description
Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate system. Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. *Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated.
Why the product was recalled
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentionally.
Distribution information
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Algeria, Armenia, Australia, Bahrain, Belarus, Belgium, Canada, China, Cyprus, Denmark, Egypt, France, Germany, Great Britain, Greece, Ireland, Italy, Albania, Bulgaria, Croatia, Czech Republic, Estonia, Hong Kong, Hungary, Iceland, India, Iran, Israel, Ivory Coast, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Montenegro, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Ukraine, Vietnam, Yemen, Argentina, Brazil, Chile, and Dominican Republic
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.