FDA recall record

Baxter Healthcare Preventative Maintenance Kits G5010007) and spare part kits recall — FDA Z-1280-2021

Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm…

Class IIclassification
Z-1280-2021official record ID
Mar 31, 2021report date

Recall record at a glance

Official recordZ-1280-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBaxter Healthcare Corporation
Report dateMar 31, 2021
Recall initiation dateJan 28, 2021
Quantity35175 kits
LocationDeerfield, IL, United States

Official product description

Preventative Maintenance Kits G5010007) and spare part kits (G5064801 and G5006203) - Product Usage: For Prismaflex Control Units, used for Continuous Renal Replacement Therapy (CRRT).

Why the product was recalled

Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.

Distribution information

Worldwide distribution - US Nationwide distribution including Puerto Rico and Guam and the countries of Albania, Algeria, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia Herzegovina, Bulgaria, Canada, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxemburg, Macedonia, Malta, Mauritius, Montenegro, Morocco, Netherlands, Nigeria, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sudan, Syria, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Yemen, Australia, New Zealand, China, Japan, Korea, Taiwan, Hong Kong, India, Thailand, Singapore, Malaysia, Vietnam, Grand Cayman, Dominican Republic, St. Kitts, Venezuela, Argentina, Chile, Barbados, Bermuda, Costa Rica, Cuba, Haiti, Guatemala, Colombia, Trinidad & Tobago, Mexico, and Brazil.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.