FDA recall record

Pentax Medical Video Duodenoscope - Product Usage: intended to be used with endoscopic recall — FDA Z-1682-2021

IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the…

Class IIclassification
Z-1682-2021official record ID
Jun 2, 2021report date

Recall record at a glance

Official recordZ-1682-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmPentax of America Inc.
Report dateJun 2, 2021
Recall initiation dateApr 21, 2021
Quantity108 units
LocationMontvale, NJ, United States

Official product description

Pentax Medical Video Duodenoscope - Product Usage: intended to be used with endoscopic devices and other ancillary equipment to provide optical visualization of (via a video monitor), and therapeutic access to, Biliary Tract via the Upper Gastrointestinal Tract, Model Number: ED34-i10T2.

Why the product was recalled

IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the associated risks with the distal end cap (OE-A63) unexpectedly becoming detached during a procedure and updated to notify users of what immediate actions should be taken in case the event occurs. This can result in unforseen events such as mucosal injury, lacerations, or bleeding of the patient. Detachment of the distal end cap (OE-A63) into the oral cavity of the patient may also result in aspiration

Distribution information

US Nationwide distribution in the states of Alabama, California, Louisiana, Maryland, Michigan, Missouri, New Jersey, North Carolina, Pennsylvania, Texas, Washington, D.C.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 27, 2025

Pentax Medical Video Processor recall — FDA Z-2329-2025

During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke…

Class IIFDA: OngoingPentax of America Inc.