FDA recall record

PENTAX Medical Video Colonoscope- Intended to provide optical visualization of recall — FDA Z-1236-2025

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become…

Class IIclassification
Z-1236-2025official record ID
Mar 5, 2025report date

Recall record at a glance

Official recordZ-1236-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPentax of America Inc.
Report dateMar 5, 2025
Recall initiation dateJan 29, 2025
Quantity152 units
LocationMontvale, NJ, United States

Official product description

PENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EC38-i20cL

Why the product was recalled

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

Distribution information

Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 27, 2025

Pentax Medical Video Processor recall — FDA Z-2329-2025

During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke…

Class IIFDA: OngoingPentax of America Inc.