Recall record at a glance
| Official record | Z-1327-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | PENTAX of America Inc. |
| Report date | May 22, 2019 |
| Recall initiation date | Dec 3, 2018 |
| Quantity | 4 devices |
| Location | Redwood City, CA, United States |
Official product description
Pentax Medical C2 CryoBalloon Standard 90 degree Catheter, Model FG 1030, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.
Why the product was recalled
Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.
Distribution information
Distribution was made to CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA. There was no government/military distribution. Foreign distribution was made to Canada, Germany, and the Netherlands.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.