FDA recall record

Medicrea International Pedicle Screws, multiple sizes, Article Nos recall — FDA Z-1472-2020

An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw…

Class IIclassification
Z-1472-2020official record ID
Mar 18, 2020report date

Recall record at a glance

Official recordZ-1472-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedicrea International
Report dateMar 18, 2020
Recall initiation dateJan 28, 2020
Quantity3529
LocationRillieux La Pape, France

Official product description

Pedicle Screws, multiple sizes, Article Nos. B36004525 B36004530 B36004535 B36004540 B36004545 B36004550 B36004555 B36005530 B36005535 B36005540 B36005545 B36005550 B36005555 B36005560 B36006530 B36006535 B36006540 B36006545 B36006550 B36006555 B36006560 B36007535 B36007540 B36007545 B36007550 B36007555 B36007560 B36008535 B36008540 B36008545 B36008550 B36008555 B36008560 B36009535 B36009540 B36009545 B36009550 B36009555 B36009560 Screws manufactured in titanium alloy Ti-6Al-4V LI conforming to ISO5832-3 and ASTM. F136 specifications. These screws are part of an internal fixation system composed of hooks, rods, plates, cross links, connection and locking devices.

Why the product was recalled

An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.

Distribution information

Worldwide distribution - US Nationwide distribution in the states of CO, FL, GA, LA, OH, and RI. International distribution to Greece and Portugal.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.