Recall record at a glance
| Official record | Z-1118-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Medicrea International |
| Report date | Feb 12, 2025 |
| Recall initiation date | Dec 3, 2024 |
| Quantity | 1,548 users |
| Location | Rillieux La Pape, France |
Official product description
UNiD" Adaptive Spine Intelligence UNiD Spine Analyzer Product Number SW3002 UDI-DI code 03613720286929 Two cloud-based software applications within UNiD Adaptive Spine Intelligence (ASI): the UNiD HUB and the UNiD Spine Analyzer. The UNiD HUB cloud-based software is a healthcare application to receive, transfer, display, store data needed for planning a spine surgery or for post-operative follow-up (patient information, X-ray image and recommendations for planning). The UNiD Spine Analyzer cloud based software is a healthcare application intended for assisting healthcare professionals in viewing and measuring images as well as planning spine surgeries.
Why the product was recalled
Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.
Distribution information
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.: Argentina Australia Belgium Brazil Bulgaria Chile Czech Republic Denmark Finland France Germany Greece Iceland Ireland Israel Italy Luxembourg Malaysia Mexico Monaco Morocco Netherlands New Zealand Poland Portugal Singapore South Korea Spain Sweden Switzerland United Kingdom
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.