FDA recall record

Philips Panorama 1.0T HFO, Magnetic Resonance (MR) System recall — FDA Z-0505-2024

Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium…

Class Iclassification
Z-0505-2024official record ID
Dec 20, 2023report date

Recall record at a glance

Official recordZ-0505-2024
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmPhilips North America LLC
Report dateDec 20, 2023
Recall initiation dateNov 17, 2023
Quantity150 US; 192 ROW
LocationCambridge, MA, United States

Official product description

Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical Electrical Systems for use as a diagnostic device.

Why the product was recalled

Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.

Distribution information

Domestic distribution to AL AZ CA CO DE FL GA IA IL IN KS KY LA MA MD MI MO NE NJ NV NY OH OK OR PA SC TN TX UT VA VT WA WI WV International distribution to Argentina Austria Canada China Colombia Croatia Czech Republic Denmark Ecuador Egypt France Germany Greece Italy Kuwait Libya Netherlands NO93 Panama Poland Portugal Qatar Russian Fed. Saint Kitts Saudi Arabia Spain Sweden Switzerland Syria Thailand T¿rkiye United Kingdom Utd.Arab.Emir. Venezuela

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.