Philips Brilliance iCT; Product Code (REF): 728306 recall — FDA Z-1315-2026

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these…

Class IIclassification
Z-1315-2026official record ID
Feb 18, 2026report date

Recall record at a glance

Official recordZ-1315-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmPhilips North America LLC
Report dateFeb 18, 2026
Recall initiation dateJan 7, 2026
Quantity47 units
LocationCambridge, MA, United States

Official product description

Brilliance iCT; Product Code (REF): 728306;

Why the product was recalled

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Distribution information

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceFeb 18, 2026

Philips Spectral CT 7500 on Rails recall — FDA Z-1318-2026

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to…

Class IIFDA: OngoingPhilips North America LLC