FDA recall record

Insulet Omnipod DASH Personal Diabetes Manager (PDM), Catalog numbers US: 18239 M/D recall — FDA Z-1553-2020

In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the…

Class IIclassification
Z-1553-2020official record ID
Apr 1, 2020report date

Recall record at a glance

Official recordZ-1553-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmInsulet Corporation
Report dateApr 1, 2020
Recall initiation dateFeb 13, 2020
Quantity32756 (US), 1420 (International)
LocationActon, MA, United States

Official product description

Omnipod DASH Personal Diabetes Manager (PDM), Catalog numbers US: 18239 M/D: PDM-USA1-D001-MG-USA1 EU: PT-000010 - M/D: INT1-D001-MG PT-000011 M/D: INT1-D001-MM UDI 10385082000009 The Omnipod DASH Personal Diabetes Manager (PDM) is a mobile, handheld Android device that is intended to communicate with the Omnipod DASH Pod (pump).

Why the product was recalled

In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the user does not exit the bolus calculator as designed or when a system alarm interrupts a bolus calculation. If the user delivers the bolus, this may lead to hypoglycemia or hyperglycemia.

Distribution information

US Nationwide distribution. Foreign distribution to Italy, Netherlands, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 8, 2026

Insulet Omnipod (Eros) Pod recall — FDA Z-2605-2026

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary…

Class IFDA: OngoingInsulet Corporation
Medical DeviceApr 22, 2026

Insulet Omnipod 5 Pods recall — FDA Z-1797-2026

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user…

Class IFDA: OngoingInsulet Corporation