Insulet Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number recall — FDA Z-2164-2020

After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract…

Class IIclassification
Z-2164-2020official record ID
Jun 10, 2020report date

Recall record at a glance

Official recordZ-2164-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmInsulet Corporation
Report dateJun 10, 2020
Recall initiation dateApr 2, 2020
Quantity400
LocationActon, MA, United States

Official product description

Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number PT000010 M/D: INT1D001MG - Product Usage: is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

Why the product was recalled

After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.

Distribution information

The products were distributed to the following foreign countries: Italy, Netherlands, UK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 8, 2026

Insulet Omnipod (Eros) Pod recall — FDA Z-2605-2026

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary…

Class IFDA: OngoingInsulet Corporation
Medical DeviceApr 22, 2026

Insulet Omnipod 5 Pods recall — FDA Z-1797-2026

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user…

Class IFDA: OngoingInsulet Corporation