FDA recall record

Insulet Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM recall — FDA Z-0423-2023

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or…

Class IIclassification
Z-0423-2023official record ID
Dec 14, 2022report date

Recall record at a glance

Official recordZ-0423-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmInsulet Corporation
Report dateDec 14, 2022
Recall initiation dateNov 14, 2022
Quantity129,086
LocationActon, MA, United States

Official product description

Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM

Why the product was recalled

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Distribution information

Domestic distribution nationwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJul 8, 2026

Insulet Omnipod (Eros) Pod recall — FDA Z-2605-2026

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary…

Class IFDA: OngoingInsulet Corporation
Medical DeviceApr 22, 2026

Insulet Omnipod 5 Pods recall — FDA Z-1797-2026

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user…

Class IFDA: OngoingInsulet Corporation