TriMed Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14 recall — FDA Z-2263-2018

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with…

Class IIclassification
Z-2263-2018official record ID
Jun 27, 2018report date

Recall record at a glance

Official recordZ-2263-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTriMed Inc.
Report dateJun 27, 2018
Recall initiation dateApr 13, 2018
Quantity323
LocationSanta Clarita, CA, United States

Official product description

Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22

Why the product was recalled

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Distribution information

U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMay 22, 2024

TriMed recall — FDA Z-1792-2024

Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.

Class IIFDA: OngoingTriMed Inc.
Medical DeviceMay 22, 2024

TriMed recall — FDA Z-1791-2024

Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.

Class IIFDA: OngoingTriMed Inc.