TriMed recall — FDA Z-1792-2024
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Overview of recall records associated with Trimed Inc., including dates, categories and classifications where available.
Classification mix: Class II: 19.
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.
The epoxy on Apex manufactured epoxy-coated instruments can pop off.
The spring on the PEG-Extender can rust due to the type of steel used to manufacture it.
The epoxy on Apex manufactured epoxy-coated instruments can pop off.
The epoxy on Apex manufactured epoxy-coated instruments can pop off.
The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and the modification may not have been fully…
The WHD/WHV impactors can rust over time due to an improper manufacturing process.
Modifications to the specification were not appropriately qualified through the company's design control process which may result in structural integrity issues of the device.
The bone screw is shorter than the intended design specification.
Two instrument holes located on the side of the plate, which act to align the saw guide are located on the wrong side of the plate
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found.…
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found.…
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found.…
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found.…
Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the…
TriMed is recalling the Semi-Tubular Plates with 10 holes (SMTP-10) because a non-locking screw passed through a hole in a SMTP-10 plate being implanted.
This page summarizes recall records in which the official source identifies Trimed Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.