FDA recall record

Medtronic Inc., Cardiac Rhythm and Heart Failure Medtronic, Viva Quadripolar Implantable recall — FDA Z-2650-2016

78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low…

Class IIclassification
Z-2650-2016official record ID
Aug 31, 2016report date

Recall record at a glance

Official recordZ-2650-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic Inc., Cardiac Rhythm and Heart Failure
Report dateAug 31, 2016
Recall initiation dateAug 12, 2016
Quantity78 (39 US, 39 OUS)
LocationMounds View, MN, United States

Official product description

Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D DTBA1D1, DTBA1D4 VIVASCRT-D DTBB1D1, DTBB1D4 Product Usage: The Medtronic Viva CRT -D dual chamber, implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT -D) is a multiprogrammable cardiac device that monitors and regulates the patients heart rate by providing single or dual chamber, rate responsive bradycardia pacing; sequential biventricular pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies.

Why the product was recalled

78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.

Distribution information

Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.