Medtronic Inc., Cardiac Rhythm and Heart Failure Amplia MRI" CRT -D SureScan", Amplia MRI" recall — FDA Z-0962-2017

During internal testing conducted as part of next generation product development, it was discovered that the AdaptivCRT (aCRT) feature of Amplia MRI and…

Class IIclassification
Z-0962-2017official record ID
Jan 18, 2017report date

Recall record at a glance

Official recordZ-0962-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic Inc., Cardiac Rhythm and Heart Failure
Report dateJan 18, 2017
Recall initiation dateDec 9, 2016
Quantity21,111 ( 13,198 US; 7,913 OUS).
LocationMounds View, MN, United States

Official product description

Amplia MRI" CRT -D SureScan", Amplia MRI" Quad CRT -D SureScan", Claria MRI" CRT -D SureScan", Claria MRI" Quad CRT -D SureScan. implantable medical devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia.

Why the product was recalled

During internal testing conducted as part of next generation product development, it was discovered that the AdaptivCRT (aCRT) feature of Amplia MRI and Claria MRI CRT -D devices may appear to be enabled when it is not active.

Distribution information

US: All states in USA including DC & PR. OUS: Australia, Finland, Japan, Saudi Arabia, Austria, France, Korea,Singapore, Bangladesh, Germany, Luxembourg, South Africa, Belgium, Greece, Malaysia, Spain, Brunei Darussalam, Guadeloupe, Netherlands, Sweden, Cayman Islands, Hong Kong, New Zealand, Switzerland, Chile, India, Norway, Thailand, Cyprus, Iran, Poland, Turkey, Denmark, Ireland, Portugal, United Arab Emirates, Dominican Republic, Italy, Reunion, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.