FDA recall record

Medtronic Curved Suction 90 9733451 EM ENT, reusable, Rx recall — FDA Z-3053-2017

The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

Class IIclassification
Z-3053-2017official record ID
Aug 30, 2017report date

Recall record at a glance

Official recordZ-3053-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMedtronic Navigation, Inc.
Report dateAug 30, 2017
Recall initiation dateJun 29, 2017
Quantity75 instruments
LocationLouisville, CO, United States

Official product description

Medtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it

Why the product was recalled

The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

Distribution information

Worldwide Distribution - US (nationwide) and Internationally to Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.