FDA recall record

Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010 recall — FDA Z-1925-2017

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when…

Class Iclassification
Z-1925-2017official record ID
May 24, 2017report date

Recall record at a glance

Official recordZ-1925-2017
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmMagellan Diagnostics, Inc.
Report dateMay 24, 2017
Recall initiation dateNov 29, 2014
Quantity49 units
LocationNorth Billerica, MA, United States

Official product description

Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010

Why the product was recalled

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Distribution information

Nationwide Distribution

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.