FDA recall record

Magellan Diagnostics The LeadCare II Blood Lead Test System relies on electrochemistry and a recall — FDA Z-0584-2026

Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with…

Class IIclassification
Z-0584-2026official record ID
Dec 3, 2025report date

Recall record at a glance

Official recordZ-0584-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMagellan Diagnostics, Inc.
Report dateDec 3, 2025
Recall initiation dateFeb 23, 2024
Quantity1,774 units
LocationNorth Billerica, MA, United States

Official product description

The LeadCare II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)

Why the product was recalled

Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.