Recall record at a glance
| Official record | Z-1245-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Tosoh Smd Inc. |
| Report date | Mar 1, 2017 |
| Recall initiation date | Dec 22, 2016 |
| Quantity | 34 |
| Location | Grove City, OH, United States |
Official product description
Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting Software is an application which serves as middle ware between the Tosoh Automated HPLC G8 analyzer and an LIS or as a stand alone data repository for the analyzer. It is designed to store and manage the received data and make reports as necessary. The application is installed on Dell All- in-One computers that do not have serial (RS232) ports only USB ports. The G8 analyzer does not have USB ports. The Key Span Adaptor connects the serial port on the G8 to the USB port on the Dell computer. Once installed the Keyspan adapter is assigned a COM port. The default for the firmware is dynamic which randomly assigns a port for the transmission of data. The port assignment can change when the power for the analyzer or computer is shut down or the cable is physically disconnected and then reconnected. The default can be changed to geographic which allows the port assignment to be static regardless of a power outage.
Why the product was recalled
Power outages causes reporting software to shutdown.
Distribution information
US Nationwide Distribution in the states of CT, MA, ME, MN, MO, NC, NE, NH, PA, SD, TX, VA, & WV.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.