FDA recall record

Tosoh Smd Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting recall — FDA Z-1245-2017

Power outages causes reporting software to shutdown.

Class IIclassification
Z-1245-2017official record ID
Mar 1, 2017report date

Recall record at a glance

Official recordZ-1245-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTosoh Smd Inc.
Report dateMar 1, 2017
Recall initiation dateDec 22, 2016
Quantity34
LocationGrove City, OH, United States

Official product description

Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting Software is an application which serves as middle ware between the Tosoh Automated HPLC G8 analyzer and an LIS or as a stand alone data repository for the analyzer. It is designed to store and manage the received data and make reports as necessary. The application is installed on Dell All- in-One computers that do not have serial (RS232) ports only USB ports. The G8 analyzer does not have USB ports. The Key Span Adaptor connects the serial port on the G8 to the USB port on the Dell computer. Once installed the Keyspan adapter is assigned a COM port. The default for the firmware is dynamic which randomly assigns a port for the transmission of data. The port assignment can change when the power for the analyzer or computer is shut down or the cable is physically disconnected and then reconnected. The default can be changed to geographic which allows the port assignment to be static regardless of a power outage.

Why the product was recalled

Power outages causes reporting software to shutdown.

Distribution information

US Nationwide Distribution in the states of CT, MA, ME, MN, MO, NC, NE, NH, PA, SD, TX, VA, & WV.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.