FDA recall record

Tosoh Smd AIA-PACK B12 Pretreatment Set- IVD for diagnostic use for pre-treatment of recall — FDA Z-0745-2020

The AIA-2000 bar code scanners depend on the quiet zone to help identify the bar code. The lack of an adequate quiet zone can cause the bar code scanner…

Class IIIclassification
Z-0745-2020official record ID
Jan 1, 2020report date

Recall record at a glance

Official recordZ-0745-2020
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmTosoh Smd Inc.
Report dateJan 1, 2020
Recall initiation dateNov 15, 2019
Quantity1012 units
LocationGrove City, OH, United States

Official product description

AIA-PACK B12 Pretreatment Set- IVD for diagnostic use for pre-treatment of calibrators/controls. Part Number: 020706

Why the product was recalled

The AIA-2000 bar code scanners depend on the quiet zone to help identify the bar code. The lack of an adequate quiet zone can cause the bar code scanner to confuse the bar codes with text or graphics resulting in the scanner ignoring a valid symbol. This results in an error in reading the barcode: Error Code 2220 Unable to read barcode on reagent bottle. In some cases, the quiet zone is too short and an error in barcode reading occurs. The problem is intermittent and not all vials (6 vials are available per set) have been reported to fail. The only risk to patients or users is a delay in results. The company recommends (optional) that customers pre-screen the vials to identify affected products. Customers can contact the company for reimbursement of any affected products. The company has implemented barcode checks as a temporary solution and will replace the barcode printer on the production line for products beginning with lot B*01898.

Distribution information

Nationwide Foreign: Columbia, Dominic Republic, Ecuador

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.